# ASTM F88: The Importance of Seal Strength Tests

[Get a Validation Plan Reviewed](https://lso-inc.com/solutions/packaging-validation/#intake-form)

FDA-Registered

Manufacturing Facility

ISO 13485

Certified Quality System

12 Cleanrooms

Class 5–8

125,000+ Sq Ft

Regulated Manufacturing

3 Facilities

CA | NH | Costa Rica

ISO 17025

Accredited Packaging Lab

/

ASTM-F88

## ASTM F88 Seal Strength Testing for Medical Device Packaging

ASTM F88 / F88M is the industry-standard test method for measuring [**seal strength**](https://lso-inc.com/medical-package-testing/services/peel-strength-testing/) in flexible sterile barrier systems. It confirms the mechanical integrity of a package seal after manufacturing, sterilization, transit simulation, and shelf-life aging.

Under [**ISO 11607**](https://lso-inc.com/medical-package-testing/standards/iso-11607/), seal strength testing is a **required component** of sterile barrier validation.

At Life Science Outsourcing (LSO), we perform ASTM F88 seal strength testing as part of medical package validation and shelf-life programs, using calibrated tensile equipment and validated procedures.

### Video Overview: ASTM F88 Explained

## What ASTM F88 Measures

ASTM F88 evaluates the **force required to peel open a sealed joint**, providing quantitative evidence of:

-   Seal uniformity
-   Adequate adhesive bonding or thermal seal strength
-   Package robustness after sterilization
-   Long-term performance following accelerated or real-time aging
-   Compliance with ISO 11607 packaging validation requirements

**Q. What does ASTM F88 measure?**  
ASTM F88 measures the _strength of the seal itself_, not the material. It determines whether a sealed joint will maintain sterile integrity during the product’s expected shelf life.

**Q. Is ASTM F88 required?**  
Yes. ASTM F88 is a recognized method for demonstrating sterile barrier performance under ISO 11607 and FDA expectations.

## Why Seal Strength Matters in Packaging Validation

Seal strength is central to sterile barrier performance. A seal that is too weak may open prematurely, while one that is too strong may tear the material instead of peeling cleanly.

F88 results support:

-   [Sterilization validation](https://lso-inc.com/validation/) (EtO, Steam, Gamma, eBeam)
-   [Accelerated](https://lso-inc.com/medical-package-testing/services/accelerated-aging/) and [real-time aging](https://lso-inc.com/medical-package-testing/services/real-time-aging/) shelf-life studies
-   Transit simulation testing ([ASTM D4169](https://lso-inc.com/medical-package-testing/standards/astm-d-standards/astm-d4169/), [ISTA](https://lso-inc.com/medical-package-testing/standards/ista/))
-   Design verification and process validation

**Q. What is an acceptable seal strength?**  
There is no universal minimum value. Acceptance criteria are set by the manufacturer based on material, seal process capability, and risk management.

**Q. Does ASTM F88 replace burst testing?**  
No. F88 measures peel strength. Burst testing (ASTM [F1140](https://lso-inc.com/medical-package-testing/standards/astm-f-standards/astm-f1140/)/F2054) measures pressure resistance.

![peel\_test](https://lso-inc.com/api/media/file/peel_test-3.jpg)

## ASTM F88 Test Techniques A, B, and C

ASTM F88 provides three test configurations. LSO documents all three  but performs **Technique A (Unsupported)**, the most commonly used method for flexible sterile barrier systems.

  
**Technique A — Unsupported**

The specimen is pulled apart without external support.

-   Most reflective of real-world opening
-   More sensitive to operator technique
-   LSO uses this method for customer studies

  
**Technique B — Supported**

The seal is supported by a plate to eliminate bending.

  
**Technique C — Restrained**

The specimen is clamped between rigid plates with an opening to expose only the seal.

These three techniques measure the same property but may produce different numerical results.

**Q. Which technique should I use?**  
For most pouches and Tyvek systems, Technique A is preferred. Technique B or C may be used when very thin films or stiff laminates require more controlled peel geometry.

## How ASTM F88 Testing Works (Step-by-Step Example)

1.  **Condition Samples**  
    Samples are conditioned to ambient laboratory conditions prior to testing.
2.  **Prepare Test Strips**  
    Cut uniform 1-inch (25 mm) wide strips perpendicular to the seal area and about 3 inches long.  The width of the strip can vary based on regulatory requirements or working area of the package.
3.  **Load Specimen**  
    Insert each end of the strip into the grips of a calibrated tensile tester.
4.  **Set Crosshead Speed**  
    LSO uses a speed of **250 mm/min** (placeholder).  Generally 200-300 mm/minute are used and can be customized.
5.  **Pull to Seal Separation**  
    The instrument measures force as the seal peels open.
6.  **Record Data**  
    Report:
7.  -   Maximum force (seal strength)
    -   Average force (seal strength)
    -   Force standard deviation
8.  **Document Failure Mode**
9.  -   Adhesive Peel (which is the ideal peel)
    -   Elongation
    -   Delamination
    -   Material tear
    -   Cohesive failure
    -   Incomplete seal (channels, seal creep, or narrow seals)

**Q. What jaw speed should I use?**  
ASTM F88 recommends 200–300 mm/min. LSO uses 250 mm/min unless otherwise specified.

**Q. How should F88 results be reported?**  
Maximum, average, the standard deviation, and sample failure mode.

## Sample Size & Seal Location Requirements

1.  One of the most overlooked elements of F88 is sample planning.  
    For compliance with ISO 11607, you must test:  
    Recommended minimums:
2.  -   **Multiple seal locations** (3 manufacturer’s seal and 1 production seal locations are most typical)
    -   **Multiple samples per location** (usually 30 or 60 depending on confidence and reliability calculations)
    -   **Samples from each production lot**
    -   **Samples from each aging time point**
3.  -   **30–60 replicates per seal location**
    -   **Time-point testing** aligned with accelerated or real-time aging
    -   **Representative sampling** for transit simulation

**Q. How many samples do I need?**  
There is no fixed ASTM rule. Industry practice is 30–60 samples per location, per time point.

**Q. Should I test across the entire seal?**  
Yes. It is best practice to test at least four locations to capture variability.

## Common Seal Failure Modes

ASTM F88 requires documenting the type of seal failure. These modes provide insight into process capability:

-   **Adhesive Peel** – Clean, expected opening
-   **Elongation** – material stretches until seal peels or material breaks
-   **Material Tear** – Film tears before seal separates
-   **Delamination** – Laminates separate; indicates material issue
-   **Cohesive Failure** – Seal adhesive releases prematurely
-   **Incomplete Seal / Channels** – Process defect requiring immediate correction

**Q. What failure modes are acceptable?**  
A clean peel is typically desired. Channels or incomplete seals, material breaks, elongation, and cohesive peels indicate failure.

![Peel Strength Testing for Medical Packaging – ASTM F88](https://lso-inc.com/api/media/file/peel-strength-testing-astm-f88-3.jpg)

## Equipment Requirements

LSO uses calibrated tensile equipment designed specifically for medical package testing with:

-   Precision load cells
-   Pneumatic or mechanical grips
-   Jaw alignment controls
-   Verified speed accuracy (200-300 mm/min)
-   Routine calibration and preventive maintenance

**Q. Can any tensile tester perform ASTM F88?**  
No. The instrument must meet speed accuracy and load accuracy defined by the standard.

## ASTM F88 and ISO 11607 Compliance

Seal strength is explicitly required under **ISO 11607-1** as part of sterile barrier system performance testing.

ASTM F88 supports:

-   **Design validation**
-   **Shelf-life validation**
-   **Package validation**
-   **Post-sterilization integrity verification**

Regulators expect F88 data in:

-   510(k) submissions
-   EU technical documentation
-   Change control justifications
-   Aging study reports

**Q. Is F88 required for ISO 11607?**  
Yes — seal strength testing is mandatory.

## Frequently Asked Questions (FAQs)

What does ASTM F88 measure?

Peel strength of a sealed joint.

Which technique does LSO use?

Technique A (Unsupported).

What jaw speed is recommended?

200–300 mm/min; LSO uses 250 mm/min.

Does ASTM F88 detect defects in the seal?

Yes — incomplete seals are considered failure modes.

Is ASTM F88 required under ISO 11607?

Yes — it is part of sterile barrier validation.

How many samples do I need?

Typically, 30–60 per location, per time point.