NewsNovember 6, 2025

![Seal\_Integrity\_600x250](https://lso-inc.com/api/media/file/Seal_Integrity_600x250-3.jpg)

Every sterile medical device package comes with an expiration date — not just for the product, but for the validated shelf life of its sterile barrier system (SBS). As that two-year mark approaches, many manufacturers assume they must start the entire validation process from scratch.

Fortunately, that’s not always necessary.

Through revalidation, you can often extend your existing packaging validation — and therefore your labeled expiration date — by up to two years or more, provided your packaging system, sealing process, and sterilization method remain unchanged.

> “Many manufacturers don’t realize their packaging validation doesn’t have to expire,” says Alan Evans, Packaging & Validation Manager at LSO. “If the system hasn’t changed, revalidation can extend your expiration date and maintain full compliance — saving significant time and cost.”

### **What Determines Your Expiration Date**

Your expiration date is defined by how long the package can maintain sterility and integrity under real-world conditions.  
This is established during initial validation using aging studies (accelerated and/or real-time) that simulate the effects of time and environmental exposure.

When you approach the end of that validated shelf life, you can generate new data — using the same standards and test methods — to confirm continued performance and safely extend the shelf life.

### **How Revalidation Extends Shelf Life**

Revalidation leverages your existing data, supplementing it with targeted testing.  
The process typically includes:

1.  **Review of Original Validation Data**
2.  **Accelerated Aging Study Extension** (per ASTM F1980)
3.  **Package Integrity Testing** (ASTM F88, F1929, F2096)
4.  **Documentation & Regulatory Justification**

If results meet all acceptance criteria, the data can support extending your labeled expiration date without a full revalidation cycle — saving time, money, and regulatory complexity.

### **Why Revalidate with LSO**

Life Science Outsourcing (LSO) provides **end-to-end packaging validation and revalidation services** — including accelerated aging, seal strength testing, and regulatory documentation support — all under one roof.

Our team ensures your packaging remains compliant with **ISO 11607-1/2**, **FDA 21 CFR part 820**, and **EU MDR** requirements while minimizing disruption to your production schedule.

**Don’t let your packaging validation expire.**  
Talk to an LSO packaging validation specialist today to explore revalidation options that can extend your shelf life — and keep your supply chain running smoothly.

## Related Articles

[Jan 29, 2026

### What the FDA’s QMSR Means for Medical Device OEMs and Manufacturing Partners

Read Article](https://lso-inc.com/news/fda-qmsr-what-it-means-for-oems/) [Dec 19, 2025

### ASTM D4169 Schedule E: Random Vibration Testing

Read Article](https://lso-inc.com/news/astm-d4169-schedule-e/) [Dec 19, 2025

### ASTM D4169 Schedule I: Low Pressure (Vacuum) Testing

Read Article](https://lso-inc.com/news/astm-d4169-schedule-i/)

## Ship faster with fewer vendors

LSO consolidates assembly, packaging, validation, and sterilization under one roof — FDA-registered, ISO 13485-certified, across three facilities. Tell us about your device and we'll map the fastest path to production.

[Request a Quote](https://lso-inc.com/contact/)

Or call us directly: [(714) 672-1090](tel:7146721090)