ISTA Standard 1A

Non-Simulation Integrity Performance Tests

About ISTA Standard 1A

ISTA 1A is for packaged-products 150 lb. (68 kg) or less, and challenges the strength and robustness of the product and package combination to withstand transport hazards. Procedure 1A is useful particularly when used as a consistent benchmark over time.

Many medical device manufacturers use ISTA Procedure 1A. This procedure was developed for a quick and easy way to evaluate a shipping configuration for drop testing and repetitive shock vibration. This procedure does not take into a account random vibration or atmospheric pre-conditioning.

Procedure

Test Procedure 1A is broken down into two categories: Vibration and Shock
  • Vibration is fixed displacement that complies with ASTM D999. Rotary or vertical linear motion of the platform is acceptable.
  • Shock testing is broken down into three test types: Drop, Incline-Impact (Conbur), and Horizontal Impact. Drop testing is a free fall test and complies with ASTM-D5276. Incline-Impact (Conbur) testing complies with ASTM-D880, and Horizontal Impact testing is in compliance with ASTM-D4003.

Benefits

When properly applied, ISTA procedure 1A will provide tangible benefits of:

  • Shortened packaged development time and confidence in product launch
  • Protection of products and profits with reduced damage and product loss
  • Economically balanced distribution costs
  • Customer satisfaction and continued business

ISTA Procedure 1A

Testing Sequence and Types:
SequenceTest TypeStandardDescription

1

Vibration

ASTM-D999

Fixed Displacement, Rotary,
Vertical Linear Motion

2

Drop

ASTM-D5276

Free Fall

Let’s work together

Life Science Outsourcing prides itself on our role as partners in bringing medical innovations to life. That partnership forms the foundation of what sets us apart.